You're a few months into EP and most of your list is single shot AF ablation. Then an SVT case comes up, the EP asks "am I early?", and do you think which one of these 40-odd channels of intracardiac electrograms (EGMs) staring back at you could they possibly be referring to, and early compared to what? Damn you gotta get better at reading EGMs you think to yourself, it's been months in the EP lab.
Where do you start, and how do you actually learn to read them?
(Click to zoom on image)
Answer:
Click to Show Answer
Why EGMs are hard to learn
An EGM is local: the tissue under a pair of electrodes, plus whatever far field leaks in. Each EGM is variably filtered with relevant vs irrelevant signals, depending on electrode size, electrode spacing, and electrode configuration.
You’re reading many channels at once, at 100 to 200 mm/s rather than 25 mm/s.
Most of us learn from whatever the list shows on the day…inevitably a lot of single shot AF ablation which rarely makes you work an EGM out.
As I wrote in the EP simulators post, exposure doesn’t automatically become understanding. Structured learning in between cases helps you build from one day to another, gathering knowledge momentum and not stagnate. We know too many people who’ve been doing EP for 10+ years, but who’s knowledge stagnated 8 years ago. In a fast paced rapidly evolving field like EP, we simply cant accept that anymore.
Learning EGMs before 3D mapping
3D systems are impressive & make many cases easier. But every point on a map is an annotated EGM. If you can’t annotate one single EGM with confidence, you can’t trust a map built from tens of thousands.
As I said in 3D maps are models. EGMs are Truth., maps can be misannotated or misrepresented, but EGMs don’t “usually” lie. In the DEEP mapping post I noted that automated annotation sometimes misses the nicest, juiciest signals, and that where to annotate isn’t settled either (there’s much interlab variation).
A basic thought, even during those boring AF cases, is to ask yourself, where is each catheter, and what should it see?
HRA: near the sinus node. Large A, small or no V.
His: across the septal tricuspid annulus. A, a small sharp His, then V.
CS: in the left AV groove. Records atrial & usually ventricular signals. Distal poles sit laterally on the left and they often record a larger V because the CS kind of dives towards the LV, (or the physician is in a ventricular CS venous branch…whoops), proximal poles near the CS ostium.
RV: apex or septum. Large V, small or no A.
Find the A, H & V for beginners
Anchor to the surface ECG.
Atrial EGMs line up with the P wave,
ventricular EGMs with the QRS.
The H sits between A & V, small & easy to lose. Can’t find it? Think catheter position before calling it absent.
Near vs far field. Local signals are sharp & high frequency. Far field is broader & blunter.
In the focal AT case from the 3D maps post, the “earliest” EGM at the RA–SVC junction likely had two components: one far field early LA, and one near field (higher frequency) RA component which is “late”. Hard to prove, but I think it’s likely.
Focal AT EGM at the RA–SVC junction, with an early far field component and a later sharp near field component.
Measuring the AH and HV intervals
AH: onset of the A on the His channel to onset of the His. AV nodal conduction. Normally 50 – 140ish. Numbers vary. Its okay if its more or less. The AH is dependent on AV nodal conduction velocity and therefore varies since AV nodal conduction velocity is heavily influenced by autonomic tone, which is influenced even by breathing or a patient’s state of sedation.
HV: onset of the His to onset of the QRS (or the earliest V on any channel, if earlier). His-Purkinje conduction. Normally 35–55 ms. A fundamental tenet of EP is that the HV doesn’t change. No matter how often you whack the patient with sedation or needles, the HV should not change.
Measure from onset every time, with calipers. A prolonged HV can change what happens to the patient next (PPM or not, Accessory pathway or not?), so tell your EP.
Step 4: Read the activation sequence (concentric vs eccentric)
Sinus rhythm: HRA first, then the His atrial EGM & proximal CS, with distal CS latest.
Ventricular pacing or a retrograde A: where is the earliest atrial EGM?
Concentric: earliest near the His / proximal CS. In keeping with the AV node, but a septal pathway can look the same.
Eccentric: earliest on the distal CS. Think left-sided pathway.
Earliest on the channels you have isn’t earliest in the heart. It’s just the earliest you can see. It’s like the person looking for their keys beneath the streetlight at night. Someone asks them, “Why are you looking underneath the streetlight?” and the person answers, “It’s the only place that I can see.”
Step 5: Bipolar vs unipolar EGMs
Bipolar: the difference between two close electrodes. Cancels much far field, good for local timing.
Unipolar: one electrode against a distant reference. Morphology doesn’t depend on wavefront direction. Local activation is the steepest negative deflection, and a QS suggests you’re at or near the origin. The catch: lots of far field. Not commonly used for 2D EP recording systems but commonly looked at in the context of VT substrate maps and 3D mapping for focal arrhythmias.
A bipolar depends on wavefront direction. If activation runs perpendicular to the line between the poles, it can be virtually zero (de Bakker). So check the unipolar, and use the bipolar to tell you which part of the unipolar is local.
Unipolar QS vs rS morphology at successful and unsuccessful focal AT sites.
Step 6: Then manoeuvres, then 3D
Once steps 1–5 are automatic, move to programmed stimulation, entrainment & the SVT manoeuvres (I unpacked one of the simpler ones in the SA-VA post). Then 3D.
What are the best resources to learn EGMs?
StatPearls, Electrophysiologic Study Interpretation: free, a good first read on catheter positions & intervals.
Josephson’s Clinical Cardiac Electrophysiology: the reference text when you want depth.
Issa, Miller & Zipes, Clinical Arrhythmology and Electrophysiology: a broad companion text.
The HRS EP laboratory standards statement: how a lab is set up & run.
Your own lab: recorded cases, senior physiologists & your EPs. Revisit the ugly tracings that didn’t fit.
If you’re committed to learning EP all the way and becoming the best you can be over your career, EP in Practice is the course for you. It’s over 100 hours across 8 programs, and Program 1 (EP fundamentals) covers reading EGMs, and Program 8 takes you into 3D mapping with the Laika 3D Mapping Simulator. If you just want the basics or a credential and then stop, it isn’t. Look elsewhere.
Other courses include the Cardiac Electrophysiology Institute of Australia or Chart Healthcare. They have various resources and courses that you might find useful depending on how deep your needs are.
My two cents:
Same order every case: catheter position, A/H/V, intervals, activation sequence. Then manoeuvres. Then 3D.
A tiny bipolar isn’t always “nothing”. Check the unipolar.
At the earliest site, ask: Is it early? Is it high frequency (local)? Is there a unipolar QS?
A 3D map is only as good as the EGMs it was built from.
Thanks for tuning in :)
Cheers
Mitch & CPiP Team
References
Koulouris S, Cascella M. Electrophysiologic Study Interpretation. [Updated 2025 Mar 28]. In: StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing. PMID 32809619. https://www.ncbi.nlm.nih.gov/books/NBK560784/
de Bakker JMT. Electrogram recording and analyzing techniques to optimize selection of target sites for ablation of cardiac arrhythmias. Pacing Clin Electrophysiol. 2019;42(12):1503-1516. doi:10.1111/pace.13817. PMID 31609005.
Haines DE, Beheiry S, Akar JG, et al. Heart Rhythm Society expert consensus statement on electrophysiology laboratory standards: process, protocols, equipment, personnel, and safety. Heart Rhythm. 2014;11(8):e9-e51. doi:10.1016/j.hrthm.2014.03.042. PMID 24814989.
Issa ZF, Miller JM, Zipes DP. Clinical Arrhythmology and Electrophysiology: A Companion to Braunwald’s Heart Disease. 4th ed. Philadelphia: Elsevier; 2023.